2014年11月23日 星期日

ECRI Institute 2015 Top 10 Health Technology Hazards

(http://bme.freebbs.tw 醫學工程論壇 同步發表)

ECRI Institute 2015 Top 10 Health Technology Hazards

2014年已到11月了,ECRI也整理出今年的的TOP10醫療科技危害(2015),作為明年的病人安全作業參考。

The List for 2015
1. Alarm Hazards: Inadequate Alarm Configuration Policies and Practices
2. Data Integrity: Incorrect or Missing Data in EHRs and Other Health IT Systems
3. Mix-Up of IV Lines Leading to Misadministration of Drugs and Solutions
4. Inadequate Reprocessing of Endoscopes and Surgical Instruments
5. Ventilator Disconnections Not Caught because of Mis-set or Missed Alarms
6. Patient-Handling Device Use Errors and Device Failures
7. “Dose Creep”: Unnoticed Variations in Diagnostic Radiation Exposures
8. Robotic Surgery: Complications due to Insufficient Training
9. Cybersecurity: Insufficient Protections for Medical Devices and Systems
10. Overwhelmed Recall and Safety-Alert Management Programs

臨床類的有的7項(警報類的事件仍是蟬聯榜首),
IT類的2項(紅色),
醫儀管理的1項(藍色),也是近年來首次提及的。



因版權問題,不能直接提供文件下載的連結,有興趣的夥伴,可以點選下列網址,簡單登記一下(有星號的部份就可以了),就可以下載文件。

https://www.ecri.org/Forms/Pages/2015hazards.aspx

以下是整理自2013~2015的比較


2015
2014
2013
1
Alarm Hazards: Inadequate Alarm Configuration Policies and Practices
各種不適當警報設定及政策造成的危害
Alarm hazards  
各種不適當警報造成的危害
Alarm hazards 
各種不適當警報造成的危害
2
Data Integrity: Incorrect or Missing Data in EHRs and Other Health IT Systems
電子健康檔案和其他醫療IT系統的資料整合失敗
Infusion pump medication errors
點滴幫浦的給藥錯誤
Medication administration errors using infusion pumps
點滴幫浦給藥量錯誤
3
Mix-Up of IV Lines Leading to Misadministration of Drugs and Solutions
點滴幫浦IV管路混雜所造的給藥錯誤
CT radiation exposures in pediatric patients
在兒科患者的CT輻射暴露
Unnecessary radiation exposures and radiation burns during diagnostic radiology procedures
非必要輻射曝露及檢查過程中輻射灼傷
4
Inadequate Reprocessing of Endoscopes and Surgical Instruments
內視鏡設備和手術器械再處理的不足
Data integrity failures in EHRs and other health IT systems
電子健康檔案和其他醫療IT系統的資料整合失敗
Patient/data mismatches in EHRs and other health IT HIT systems
電子病歷及其他健康資訊資料錯誤
5
Ventilator Disconnections Not Caught because of Mis-set or Missed Alarms
呼吸器管路不合或忽視警報所造的管路脫落
Occupational radiation hazards in hybrid ORs
混合式手術室的職業輻射危害
Interoperability failures with medical devices and health  IT systems
醫儀與資訊系統連線管理疏忽
6
Patient-Handling Device Use Errors and Device Failures
協助安置病人設備使用錯誤及故障
Inadequate reprocessing of endoscopic devices and surgical instruments
內視鏡設備和手術器械再處理的不足
Air embolism hazards
空氣栓塞危害
7
“Dose Creep”: Unnoticed Variations in Diagnostic Radiation Exposures
未注意診斷時輻射的變化所產生的劑量問題
Neglecting change management for networked devices and systems
忽略聯網設備和系統的變更管理
Inattention to the needs of pediatric patients when using "adult" techniques
用成人治療技術未注意小兒病患的差別需求
8
Robotic Surgery: Complications due to Insufficient Training
由於訓練不足所導致的機器人手術併發症
Risks to pediatric patients from "adult" technologies
使用"成人"技術時導致的兒科患者的風險
Inadequate reprocessing of endoscopic devices and surgical instruments
軟式內視鏡及手術器械的交叉污染
9
Cybersecurity: Insufficient Protections for Medical Devices and Systems
醫療儀器上的資安問題
Robotic surgery complications due to insufficient training
由於訓練不足所導致的機器人手術併發症
Caregiver distractions from smartphones and  other mobile devices
智慧型手機或行動裝置造成醫護人員分心
10
Overwhelmed Recall and Safety-Alert Management Programs
過多的醫儀召修及安全警示管理要求
Retained Devices and
Un-retrieved Fragments
遺留的物件和無法恢復的碎片
Surgical Fires
手術室火災


2014年11月11日 星期二

Sharplan 20C CO2雷射輸出時有時無

(http://bme.freebbs.tw 醫學工程論壇同步發表)
名稱: Sharplan 20C CO2
故障情形:雷射輸出時有時無
故障原因:腳踏開關故障

調了許久,也許是接觸不良,好像都沒那麼根本解決,廠商報價新的踏板,不便宜哦,

拆開看,裡面有一個部份很可疑,包起來又灌膠,會不會裡面有藏電路板,果真如此,那不買原廠的也不行了,要如何確認,有點傷腦筋,只好拿出絶招....


這張圖片....沒錯,就是照 X光。只要是很難拆、轉向很難確認、可疑的...照X光是一個很有效的解決方法喔。

一照之下,原來只有接線,沒電路板,簡單的腳踏開關,裡面是二個小微動開關,靠一隻銅片要同時作動,既然沒電路板,那就單純多了,重新調整銅片的曲度,試用看看囉。 



2014年11月6日 星期四

彎針變直針

(http://bme.freebbs.tw 醫學工程論壇同步發表)
名稱: 彎針變直針